| Model No. | YSD186E-Vet | 
| Infusion Mode: | Drip-rate and Volumetric | 
| Pumping Mechanism: | Peristaltic finger pump | 
| IV Sets: | Standard IV sets of 10, 15, 20 and 60d/mL | 
| Optional Rate Unit: | d/min, mL/h and volume/time | 
| FlowRate Range: | 10, 15, 20 d/mL IV sets: 1.0~1200 mL/h | 
| 60 d/mL IV set: 1.0~150.0 mL/h | |
| Increment: | 0.1mL/ h | 
| DripRate Range: | 10 d/mL IV set: 1.0~200 d/min | 
| 15 d/mL IV set: 1.0~300 d/min | |
| 20 d/mL IV set: 1.0~400 d/min | |
| 60 d/mL IV set: 1.0~150 d/min | |
| Volume Limit: | 1~9999 mL | 
| Volume Infused Display: | 0~9999 mL & >9999 mL | 
| Flow Rate Accuracy: | ≤ ± 5% (after calibration) | 
| Drip Rate Accuracy: | ≤ ± 1.5% | 
| Purge Rate: | 600 mL/h | 
| KVO Rate: | 1 mL/h | 
| Bolus Mode: | Bolus rate adjustable | 
| Occlusion Pressure: | Pressure detection device with 8 sensitivities setting, 40 ~ 140KPa. | 
| Air-in-line Detection: | Air-in-line detection device with 8 sensitivities setting, 50 ~ 1000μL. | 
| IV Set Management & Flow Rate Calibration: | Up to 20 IV sets of customers' brand can be added, stored and deleted in memory, convenient for operation after calibration | 
| Display: | Blue LCD with high brightness | 
| Power Source: | AC: 100V ~ 240V, 50Hz/60Hz, ≤ 25VA | 
| Battery: | Rechargeable Ni-MH battery: DC12V,2000mAh, life: ≥ 4 hours at 25mL/h | 
| Classification: | ClassII, CF, IPX4 | 
| Dimensions and N.W.: | 150mmx 150mm x 235mm; 2.3kg | 
| Warranty: | Two years |